The Ugaiya Mini rapid biological reader is suitable for reading the sterilization results of the steam sterilizer
/ VHP / EO. It can monitor the survival of Bacillus Stearothermophilus spores in the rapid biological indicator.
It must be used with a specially recommended indicator.
The reader is used to cooperate with constant temperature incubation indicators at 58 °C / 37 °C
(the temperature can be adjusted according to the different requirements of the biological indicator). The final negative results
/ incubation positives were obtained automatically by reading the fluorescence changes of the biological indicator
special in VH2O2 20min / stem20min / 60min / 180min / EO240min.
In the design of the mini rapid biological reader, the application requirements of visual interpretation are considered
of the PH color. The reader is allowed to continuously incubate the special indicator at 58 °C / 37 °C for 24-48 hours.
Final negative / positive incubation results can be determined by visual interpretation of the color change
PH.
Compared with the visual interpretation of PH color change, the fluorescence intensity interpretation method
it has the advantages of high sensitivity, fast response and small error.
The equipment does not support the sterilization result test of other long-term incubation indicators except
indicators manufactured by the manufacturer
Declaration of conformity, Warranty certificate, ISO